CHESTERLAND OH - The new Series 3 DT80 range data loggers from dataTaker are intelligent monitoring solutions used the world over in almost every imaginable field and application. CAS DataLoggers Application Specialists have put together this how-to guide so users can check the calibration of their DT80 range data loggers. Just complete the 5-step procedure described below--if the logger fails any of these tests, then it should be returned to dataTaker or your local distributor for calibration.

All measurements should be conducted in a constant temperature-controlled environment (i.e. 2C) during the course of testing. This prevents temperature drift effects of the reference voltage and reference resistances. The limits shown in the sample report assume that the test environment is between 5C and 40C. For the same reason, the datalogger needs time to "soak" in a temperature-controlled environment before calibration can be performed. The unit must be powered up and left to sit at the test temperature for a minimum of 4 hours before calibration.

Also note that before conducting any test, the logger must first be earthed. Do this by connecting an earth wire to the earth point on either side of the data logger, or for older models without the earth points, use the negative ('-') input of the power terminal block immediately below the Ethernet port on the left side of the datalogger. All tests assume that the operator is familiar with using the dataTaker command language and a suitable terminal application such as DeTransfer.

Equipment required:
1x Programmable Precision voltage source*
1x 100 W 0.05% 5ppm or better
1x 1000 W 0.05% 5ppm or better
*The high-voltage range of the logger has an input impedance of approximately 110 kW. Ensure that the voltage source has low output impedance or the test results will be invalid.

1. Perform a self-test of the datalogger by issuing the TEST command and observing the result. The logger will return various test results and an overall Pass/Fail result.

2. Now perform a low-voltage range test. Using a precision voltage source, apply these known voltages to the datalogger:
+25mV, +250mV and +2500mV
-25mV, -250mV and -2500mV
Measure these voltages using the following wiring configuration and dataTaker commands: Wiring Configuration
Connect the applied voltage between the positive ('+') and negative ('-') terminals of analog channel 1.
dataTaker Commands
RA1S 1V(FF6)
Record and check results against the limits shown in the sample report.

3. Now perform a high-voltage range test: Using a precision voltage source, apply these known voltages to the logger:
+0.25V, +2.5V and +25V
-0.25V, -2.5V and -25V
Measure these voltages using the following wiring configuration and dataTaker commands. Wiring Configuration
Connect the applied voltage between the '+' and '-' terminals of analog channel 1.
dataTaker Commands
RA1S 1HV(FF6)
Record and check results against the limits shown in the sample report.

4. Perform a low-resistance range test. Apply the following known resistance to the logger:
100 W (0.05% 5ppm or better)
This test will use the "II" current excitation (2.5mA).
Measure the resistance using the following wiring configuration and dataTaker commands: Wiring Configuration
Connect the known resistance using a 4-wire configuration to analog channel1.
dataTaker Commands
RA1S 1R(4W,II,FF3)
Record and check results against the limits shown in the sample report. Move the connector to Channel 2 and repeat. Modify the command to reflect the new channel under test.
RA1S 2R(4W, ll,FF3)
Continue the above process, moving the connector to each successive channel of the dataTaker until all channels have been validated.

5. Finally, perform a high-resistance range test. Apply the following known resistances to the logger:
100 W (0.05% 5ppm or better)
1000 W (0.05% 5ppm or better)
This test will use the "I" current excitation (213A).
Measure the resistances using the following wiring configuration and dataTaker commands:
Wiring Configuration
Connect the known resistances using a 4-wire configuration to analog channel 1. dataTaker Commands
RA1S 1R(4W,I,FF3)
Record and check results against the limits shown in the sample report. Move the connector to Channel 2 and repeat. Modify the command to reflect the new channel under test.
RA1S 2R(4W, l,FF3)
Continue the above process, moving the connector to each successive channel of the dataTaker until all channels have been validated.

For further information on the dataTaker Series 3 DT80 range of intelligent data loggers, other products in the rugged dataTaker line, or to find the ideal solution for your application-specific needs, contact a CAS Data Logger Applications Analyst at (800) 956-4437 or visit the website at www.DataLoggerInc.com.

Contact Information: CAS DataLoggers, Inc. 12628 Chillicothe Road Chesterland, Ohio 44026 (440) 729-2570 (800) 956-4437 sales@dataloggerinc.com

http://www.dataloggerinc.com

Print | 
Email |  Comment   Share  
Contacts: View detailed contact information.


 

Post a comment about this story

Name:
E-mail:
(your e-mail address will not be posted)
Comment title:
Comment:
To submit comment, enter the security code shown below and press 'Post Comment'.
 



 See related product stories
More .....
 See more product news in:
Test and Measuring Instruments
 More New Product News from this company:
Thermistor Probe suits medical monitoring applications.
Data Acquisition System provides machine fault diagnostics.
Remote Temperature Recording Kit verifies thermal sterilization.
Data Acquisition System targets universities and research.
Temperature and Humidity Data Logger features graphic screen.
More ....
| Featured Manufacturing Jobs
 Other News from this company:
Managing Remote Water Pumps Using a Modem Data Logger
Remote Flood Monitoring System Sends Data via GPRS
Recording Closed-Loop Control System Cycles for HVAC Savings
CAS DataLoggers is Attending CSIA 2014
Vaccine Safety: What's It Worth?
More ....
 Tools for you
Watch Company 
View Company Profile
Company web site
More news from this company
E-mail this story to a friend
Save Story
Search for suppliers of
Data Loggers & Data Logging Equipment
Join the forum discussion at:
Wired In


Home  |  My ThomasNet News®  |  Industry Market Trends®  |  Submit Release  |  Advertise  |  Contact News  |  About Us
Brought to you by Thomasnet.com        Browse ThomasNet Directory

Copyright © 2014 Thomas Publishing Company. All Rights Reserved.
Terms of Use - Privacy Policy



Error close

Please enter a valid email address