New FCX80 Free Chlorine Analyzer Operates Over a Temperature Range from 0 to 50 degree C

Press Release Summary:

  • Designed to monitor free chlorine in drinking water, rinse water, cooling water or other fresh water samples from 0.05 to 20 ppm chlorine
  • Features a plug and play design that incorporates constant head flow control device, pH sensor, free chlorine sensor and analyzer/transmitter
  • 24-VDC powered and comes with a 128-x-64 pixel LCD display with black graphics

Original Press Release:

New FM/ATEX Approved FCX80 Free Chlorine Analyzer Designed for Hazardous Area Applications

Anaheim, CA — April 15, 2019 — With its advanced plumb and play design, economical reagent-less sensors and rugged construction, the new FCX80 Free Chlorine Analyzer from Electro-Chemical Devices is FM/ATEX/IECEx approved for hazardous area water and wastewater treatment processes in municipal and industrial plants.

The panel mounted FCX80 is a ready-to-use free chlorine analyzer intended for hazardous locations. It is designed to monitor free chlorine in drinking water, rinse water, cooling water or other fresh water samples from 0.05 to 20 ppm chlorine as the standard range or 0.01-5.00 ppm with the optional low range sensor. The FCX80 is compliant with U.S. EPA Method 334.0 for measuring drinking water.

With its FM/ATEX/IECEx approvals, the FCX80 Analyzer’s design is highly resistant to combustible gases that can be present in water treatment areas, which helps prevent explosions and fires that could jeopardize employee safety, plant equipment and the surrounding area. Such gases have been responsible for serious accidents causing loss of life and catastrophic facility damage.

ECD’s highly intelligent, reagent-free S88 Free Chlorine Sensor includes digital communication capabilities and features amperometric sensing based on polarographic technology relying on gold/silver materials and a PTFE membrane. It operates over a wide temperature range from 0 to 50°C (32 to 122°F) for use in extreme environments. Response time is T90 in 2 minutes. The only maintenance required is the replacement of its economical electrolyte solution at intervals of up to 12 months.

Free chlorine for disinfection applications exists in solution as a pH dependent ratio of hypochlorous acid (~100% at pH 5) and hypochlorite ion (~100% at pH 10). Chlorine has been a popular disinfectant for water treatment to ensure public health. Disease causing organisms such as Cholera, Hepatitis A, Typhoid Fever and Dysentery are eliminated with Chlorine, along with slime bacteria, algae and molds that can grow in water storage facilities such as reservoirs and storage tanks.

The FCX80 Analyzer features a plug and play design that incorporates a constant head flow control device, a pH sensor, the free chlorine sensor and the analyzer/transmitter all conveniently mounted on a PVC panel. Technicians simply connect the sample and drain lines, connect the power and outputs and the FCX80 Analyzer is ready to use in less than 30 minutes.

The FCX80 Analyzer is 24-VDC powered and comes with a 128-x-64 pixel LCD display with black graphics and/or type on a grey or white background with an LED backlight. This easy-to-read display allows parameters to be graphically displayed with a user defined line, bar or gauge style graph. 

For ease of plant management, the FCX80 Analyzer is compatible with supervisory control and data acquisition (SCADA) systems or large distributed control systems (DCS). The standard configuration features two 4-20 mA outputs, three single-pull/double-throw (SPDT) alarm relays and a MODBUS remote telemetry unit (RTU).

About Electro-Chemical Devices

Electro-Chemical Devices (ECD) is a manufacturer of liquid analytical process instrumentation. Founded in 1977, ECD has been in business for over 30 years and has built its industry leading reputation by its commitment to customers through this basic approach: Provide quality products, develop technical innovations and provide responsive, knowledgeable service.

ECD Company Contact:

Eric Kim

ECD Address: 1500 N. Kellogg Dr., Anaheim, CA 92807 USA

Web: www.ecdi.com

Email: sales@ecdi.com

Tel: 800-729-1333 Tel: 714-695-0051

Fax: 714-695-0057

Editor’s Contact:

Randy Brown

Tel: 909-223-6700

Email brownr@rbmarketing.com