IR Analyzer measures antioxidant inhibitor in transformer oils.

Press Release Summary:



With 2.7 micron filter specific for OH absorbance band and 0.2 mm path length sealed cell, InfraCal Filtometer can accurately measure antioxidant inhibitor on site from 5% down to 0.05% in less than 1 minute. Internal calibration table and operator interface allows for analysis by non-technical personnel. Unlike FT-IR spectrometers, fixed filter infrared analyzer has no moving parts and insignificant optical air path, making it portable and suitable for field environments.



Original Press Release:



Portable Infrared Analyzer Measures Antioxidant Inhibitor in Transformer Oils in Less than One Minute



Monitoring the degradation of the antioxidant in transformer insulating oil can result in substantial savings by extending the service life of the oil and by slowing down the transformer aging process. 2,6 ditertiary-bytyl paracresol (DBPC) and 2,6-di-tert-Butyl Phenol (DBP) are common inhibitors that prevent oxidation of the electrical insulating oil in transformers. ASTM Method 2668 specifies measuring DBPC and DBP concentrations by infrared analysis which has typically been done using an FT-IR spectrometer. However, regular testing is typically not done as often as it should because of the time and expense to transport samples to a laboratory. Due to its complexity, cost and environmental requirements, an FT-IR spectrometer is usually impractical to move out of the laboratory into the field.

The InfraCal Filtometer from Wilks Enterprise, with a 2.7 micron filter specific for the OH absorbance band (common to DBPC and DBP) and a 0.2 mm pathlength sealed cell can accurately measure the inhibitor on site from 5% down to 0.05% in less than one minute. It is an inexpensive, easy-to-use field analyzer. The InfraCal Filtometer enables more frequent testing of the antioxidant inhibitors and permits measurements to be made on-site with a substantial cost savings over laboratory analysis.

The InfraCal Filtometer is a fixed filter infrared analyzer that, unlike an FT-IR spectrometer, has no moving parts and an insignificant optical air path. This makes it portable, rugged and suitable for use in a field environment. An internal calibration table and easy-to-use operator interface allows for analysis by non technical personnel. DBPC and DBP have a characteristic infrared absorption band at 2.7 microns (3650 cm-1). As the inhibitor is depleted, the infrared absorbance decreases. An internal calibration table in the InfraCal Filtometer allows for a digital readout in % DBPC or DBP.

For further information, contact: Wilks Enterprise, Inc., 25 Van Zant Street, Ste. 8F, E. Norwalk, CT 06855 USA, TEL: 203-855-9136, FAX: 203-838-9868, Email: info@wilksir.com; or visit our Web Site -- wilksir.com, or download information directly at wilksir.com/pdf/AntioxidantinTransformeroil.pdf

About Wilks Enterprise, Inc.

Headquartered in E. Norwalk, Connecticut, Wilks Enterprise specializes in easy-to-use, portable mid-infrared analyzers for specific measurement applications. All Wilks analyzers have been specifically designed for on-site use by non-technical personnel and are used for a wide variety of applications in the biofuels, environmental, petrochemical, quality control, and manufacturing industries. These portable analyzers enable analytical measurements to be made in the field, helping to eliminate the wait for off-site lab results.